What is the Viscosity Proportional To in a Capillary Viscometer?

Published in Fluid Viscosity Measurement 3 mins read

In a capillary viscometer, the viscosity of a fluid is primarily and directly proportional to the pressure drop required to maintain a specific laminar flow rate through the capillary tube.

Understanding Viscosity's Direct Proportionality

This fundamental relationship is derived from the principles of fluid dynamics, particularly as described by the Hagen-Poiseuille equation. According to this equation:

"According to the Hagen–Poiseuille equation, the pressure drop of laminar flows in a capillary at a given flow rate is proportional to the viscosity of the fluid."

This means that for a fixed capillary tube (constant length and radius) and a consistent flow rate, a greater pressure difference is necessary to push a more viscous fluid through the tube. Conversely, if you observe a certain pressure drop, a higher pressure drop implies a higher viscosity for the fluid being tested under those conditions.

Key Factors Related to Viscosity Measurement in Capillary Viscometers

While pressure drop is the direct proportionality highlighted by the Hagen-Poiseuille equation for a given flow rate, practical capillary viscometry often measures other parameters that are intrinsically linked to viscosity.

  • Pressure Drop ($\Delta P$): This is the most direct proportionality based on the Hagen-Poiseuille equation. For a specific capillary design and a constant flow rate ($Q$), dynamic viscosity ($\eta$) is directly proportional to the pressure drop ($\eta \propto \Delta P$).
  • Flow Time ($t$) and Fluid Density ($\rho$): In many common capillary viscometers (such as the Ostwald or Ubbelohde types), the driving pressure is generated by the hydrostatic head of the fluid itself. This hydrostatic pressure is proportional to the fluid's density ($\rho$). The flow rate ($Q$) is determined by measuring the time ($t$) it takes for a fixed volume ($V$) of fluid to pass through the capillary ($Q = V/t$). Consequently, the dynamic viscosity ($\eta$) is found to be proportional to the product of the flow time and the fluid's density ($\eta \propto t \cdot \rho$). Similarly, the kinematic viscosity ($\nu = \eta / \rho$) is directly proportional to the flow time ($t$) ($\nu \propto t$).
  • Capillary Geometry (Length $L$ and Radius $R$): The physical dimensions of the capillary tube are crucial constants for a specific viscometer. The resistance to flow is directly proportional to the length and inversely proportional to the fourth power of the radius ($R^4$). These constant factors are typically incorporated into a "viscometer constant" during calibration.

Practical Application in Capillary Viscometry

Capillary viscometers, such as the widely used Ostwald viscometer, leverage these proportionalities to accurately measure fluid viscosity:

  • Measuring Flow Time: Instead of directly measuring pressure drop, these instruments typically measure the time it takes for a known volume of fluid to flow through a precisely calibrated capillary. By comparing this flow time to that of a reference fluid with known viscosity, or by using a pre-calibrated viscometer constant, the unknown viscosity can be determined.
  • Controlling Conditions: For accurate and reproducible results, it is essential to:
    • Maintain a constant and precise temperature, as viscosity is highly sensitive to temperature changes.
    • Ensure laminar flow conditions, as turbulent flow would invalidate the Hagen-Poiseuille relationship.
    • Thoroughly clean the viscometer between measurements.